Intra- and inter-rater reliability of muscle thickness measurement of the tibialis anterior using different inward pressures
Physical Therapy Rehabilitation Science 2019;8:218-24
Published online December 30, 2019
© 2019 Korean Academy of Physical Therapy Rehabilitation Science.

Seong-Joo Leea, Ji Young Lima,b, Chang-Hyung Leec, Dae-Sung Parkd

aDepartment of Physical Therapy, General Graduate School of Medical Sciences, Konyang University, Daejeon, Republic of Korea
bResearch Institute for Future Medicine, Samsung Medical Center, Seoul, Republic of Korea
cRehabilitation Medicine, Pusan National University School of Medicine, Research Institute for the Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Yangsan, Republic of Korea
dDepartment of Physical Therapy, College of Medical Science, Konyang University, Daejeon, Republic of Korea
Correspondence to: Dae-Sung Park (ORCID https://orcid.org/0000-0003-4258-0878)
Department of Physical Therapy, College of Medical Sciences, Konyang University, 158 Gwanjeodong-ro, Seo-gu, Daejeon 35365, Republic of Korea
Tel: 82-42-600-6419 Fax: 82-42-600-6565 E-mail: daeric@konyang.ac.kr
Received October 4, 2019; Revised October 28, 2019; Accepted October 29, 2019.
cc This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract

Objective

This reliability study examined the effects of applying varying induced inward pressures using a transducer placed at 0° neutral ankle position (NEU) and 15° ankle dorsiflexion (DF) on tibialis anterior (TA) muscle thickness using a custom-made device with a force indicator during rehabilitative ultrasound imaging.

Design

Cross-sectional study.

Methods

Twenty-four healthy subjects were recruited in this study. Two examiners measured the muscle thickness of the TA at 0° NEU and 15° DF in 3 conditions of inward pressures (1.0 N, 2.0 N, and 4.0 N) using a custom-made holder. The muscle thickness was measured three times for each of the conditions arranged in random order. For intra- and inter-rater reliability, the intra-class correlation coefficients (ICCs) with 95% confidence intervals, standard error of measurement, minimal detectable change, and coefficient of variation were analyzed. One-way repeated measures analysis of variance was conducted for investigating changes of TA muscle thickness according to the inward pressures of the transducers.

Results

The intra-rater reliability of TA muscle thickness measurement was excellent (ICC3,1: 0.92-0.96) for all conditions (at both ankle joint angles per varying inward pressure). Likewise, the inter-rater reliability of TA muscle thickness measurement was excellent (ICC2,1: 0.89-0.97) under same conditions. The mean of TA thickness showed the trend of decreasing significantly with increased inward pressures at all ankle joint angles (p<0.05).

Conclusions

Use of this custom-made device with a force indicator is useful to accomplish the high intra- and inter-rater reliability of TA muscle thickness measurement at both ankle joint angles in reducing the measurement error.

Keywords : Anterior tibial muscle, Lower extremity, Reproducibility of results, Ultrasonography
Introduction

Muscle thickness is the most common parameter among the muscle architecture parameters and is commonly obtained as an ultrasound image [1], and it can also be used to predict muscle size obtained from magnetic resonance imaging [2]. Moreover, muscle thickness partially contributes to muscle strength and the degree of functional impairment in neurological disorder [3]. Muscle thickness have been measured using real-time ultrasound imaging, which is a noninvasive and safe method [4]. In obtaining the skeletal muscle architecture of diverse muscles such as the core, upper limb, and lower limb muscles, its reliability and validity have been demonstrated [5-8].

Muscle imaging is considered to be a quantitative evaluation tool and a anticipated follow-up tool [9]. Notwithstanding, it is generally accepted that muscle imaging is dependent upon the skill of the operator, and measurement errors may occur [10]. For this reason, measurement errors should be reduced and an examination for the reliability of measuring muscle architecture parameters is needed. In terms of its operator dependency, one of the primary factors that can affect the imaging process is the inward pressure of the transducer [11]. A previous study [12] clarified that inward-pressures did have an effect on the thickness of the transversus abdominis, internal oblique, and external oblique muscles in various four conditions (0.5 N, 1.0 N, 2.0 N, and 4.0 N) on. Likewise, muscle thickness of the lumbar multifidus and external oblique muscles showed similar results in two inward pressure conditions (0.1 N and 2.0 N) [13].

Therefore, in order to reduce the influence of these inward pressures on muscle thickness measurements, some studies [12,13] have developed a custom-made holder that is hands-free, and have examined its reliability and the influence of different inward pressures of the transducer for imaging of the lateral abdominal and lumbar multifidus muscles. On the other hand, a recent study [14] on patients with chronic back pain used the transducer force device and real-time on-screen display of transducer force (N), inclination and roll (°) via a LabVIEW on a laptop. Some devices are not only impractical in real clinical settings operated by health care professionals but most studies have also focused on mainly the abdominal muscles. There are a lack of studies that have investigated the use of a custom- made holder and inward pressure indicator on the tibialis anterior (TA) muscle, which contributes to walking function.

In this regard, the TA muscle plays an important role in ankle dorsiflexion (DF) where sufficient activation prevents the appearance of foot drop during the swing phase of gait. TA muscle dysfunction has been mainly shown in patients with neurological injury, such as stroke [15] and cerebral palsy [16]. For functional ambulation, the ankle joint has to have at least 10° to a maximum of 20° of DF. A previous study reported that in the swing phase of gait, the greater the maximum DF range, the larger the TA muscle thickness [16]. When performing maximum voluntary isometric contraction at 15° of DF, the pennation angle of the TA muscle dramatically increased as well as the maximal torque [17]. TA muscle thickness during maximal DF was greater than that during the resting position in stroke survivors with impaired muscle properties of the lower extremity. Also, TA muscle thickness of the paretic side in persons with stroke showed a significant difference compared to the non-paretic side in 0° neutral ankle position (NEU) and 15° DF [6]. Comprehensively, this may indicate that it is clinically crucial to measure TA muscle thickness in 15° DF, which infers on functional properties such as gait.

Therefore, a hand-held transducer device connected with an indicator showing inward pressure in real-time was utilized with consideration of the effect of inward pressure on TA muscle thickness and usability in clinical settings. The purposes of this study were (1) to examine the usability of a novel hand-held method in healthy subjects, (2) to apply this method for obtaining thickness images of the TA muscles in 0° NEU and 15° DF positions at varying inward pressures of the transducer, (3) to analyze the intra- and inter-rater reliability of the measurements using the hand-held transducer with an indicator to adjust the inward pressures of the transducer, and (4) to compare changes in TA muscle thickness according to the inward pressures of transducers.

Methods

Inclusion and exclusion criteria

Subjects were recruited from Konyang University student population from Daejeon. Subjects who had a history of fracture, operation, inflammatory disease, or neurological injury in the ankle and knee joints were excluded. All participants submitted their informed consent after understanding the entire procedure of the study, which was approved by the Institutional Review Board of Konyang University (IRB No. KYU 2016-079).

Experimental equipment and procedure

For accurate measurements of ankle joint angles, an isokinetic dynamometer (HUMAC Norm; CSMi, Stoughton, MA, USA) was set at two different ranges in resting position (i.e. 0° NEU and 15° DF). Subjects were instructed to sit upright in the assessment chair and to fully extend the knee. After the ankle joint angles of the subjects were properly established, the primary examiner marked the position of the ultrasound transducer according to the reference of the scanning site that was located at 20% of the distance from the head of the fibula to the tip of the lateral malleolus. Once an appropriate transducer position was confirmed, both examiners obtained images of the right TA using real-time B- mode ultrasound imaging (MicrUS-EXT-1H; Telemed, Vilnius, Lithuania) with a 7.5 MHz linear transducer. For obtaining images in different inward pressure of the transducer (1.0 N, 2.0 N, and 4.0 N, respectively), a custom-made device with a one-axis digital load cell sensor (CSBA-10LS; Curiotech Inc., Seoul, Korea) (Figure 1A) was utilized and enabled the ability to check the inward pressure of the transducer in real-time with an indicator (CI-2001A; CAS Korea, Seoul, Korea) (Figure 1B). Although the two physical therapists had more than three years of experience in measuring muscles thickness using rehabilitative ultrasound imaging, they were additionally trained by a specialist (PhD) in rehabilitative ultrasound imaging for two days on collecting an exact image of the TA.

To check for intra-rater reliability, examiner A conducted measurements on one day and then again on the following day. To find out the inter-rater reliability, primary examiner A captured an image first, and then examiner B conducted an in an equal manner, one hour later on the same day. The subjects participated in a total of three sessions. The examiner measured the muscle thickness three times under each condition (three inward pressures [1.0 N, 2.0 N, and 4.0 N] * two ankle positions [0° NEU and 15° DF]), and eighteen images were produced per session. In total, fifty-four images were obtained per participant. The order of conditions was randomized to decrease measurement bias.

To minimize the measurement bias, another examiner who had not participated in image collection had analyzed the TA muscle thickness using the Image J software (Bethesda, MD, USA). The TA muscle thickness was defined as the liner distance between the superficial aponeurosis and the deep aponeurosis upon the center longitudinal line of the captured images.

Sample size calculation and data analysis

Based on a priori power analysis (G*Power Software, version 3.1.9.2; Heinrich-Heine-Universität, Düsseldorf, Germany), the number of group (single-group), number of measurements, and 90% of statistical power at an α level of 0.05 for the TA muscle thickness value was determined. A sample size of 24 was sufficient to optimize for repeated measures within factors.

Descriptive statistics were used for the general characteristics of the subjects. For primary analyses, based on the studies [12,13] related to inward pressure, three values of TA muscle thickness per each condition was averaged. The reliabilities of each TA thickness variable (0° NEU and 15° DF in different inward pressures of 1.0 N, 2.0 N, and 4.0 N, respectively) was estimated using intraclass correlation coefficients (ICCs) values of relative reliability with 95% confidence intervals. Intra-rater reliability was calculated based on the single-rater, absolute-agreement, 2-way mixed-effects model (ICC3,1). Inter-rater reliability was calculated based on single-rating, absolute-agreement, 2-way random effects model (ICC2,1) [18].

For investigating absolute-agreement of the measurements, the standard error of the measurement (SEM) and minimal detectable change (MDC) values were estimated by using the formula: SEM; SD×1ICC2 and MDC; SEM×1.96×2 [19]. Additionally, the coefficient of variation (CV) was computed to examine the relative measurement error. The Bland and Altman plot [20] was used to express the agreement according to each inward pressure between session 1 and session 2 of intra-rater reliability at 0° NEU and 15° DF.

To analyze for statistically significant changes of TA muscle thickness according to the inward pressures of the transducer in each ankle joint angle, a one-way repeated measures analysis of variance was conducted for within subjects’ design. Significance level of 0.05 was chosen for testing statistical significance. All statistical analyses were conducted using PASW Statistics for Windows, Version 18.0 (SPSS Inc., Chicago, IL, USA).

Results

Participant characteristics

For this single-group repeated-measures reliability study, a total of twenty-four healthy subjects (16 women and 8 men) were enrolled. Mean (SD) values of their general characteristics were 21.4 (1.4) years for age, 164.8 (8.3) cm for height, and 59.9 (11.6) kg for weight, respectively.

Reliability

The intra-rater ICC values ranged from 0.92 to 0.96 for each of the three conditions at both ankle joint angles (Table 1). ICC values at 15° DF for the three conditions were the same or greater than at the 0° NEU position. In absolute reliability of measurements, ranges of SEM and MDC were from 0.31 to 0.48 and from 0.86 to 1.26 respectively with lower SEM and MDC for 15° DF than 0° NEU. Moreover, range of CV was from 0.09 to 0.16 between session 1 and session 2. The Bland and Altman plots [20] for intra-rater reliability showed good agreement at both ankle joint angles as shown in Figure 2.

For inter-rater reliability, ICCs were shown to be at least 0.89 to maximum 0.97 for each of the three conditions at all ankle joint angles (Table 2). The ICCs values for 15° of DF (0.97) showed high levels of consistency compared to 0° NEU (ICCs: 0.89-0.91). In absolute reliability of measurements, ranges of SEM and MDC were from 0.23 to 0.56 and from 0.64 to 1.56 respectively with lower SEM and MDC for 15° DF than 0° NEU, along with the intra-rater reliability. Also, the results ranged from 0.06 to 0.17, which was within 10 % of the measures values.

Difference of TA muscle thickness at the three inward pressures

There were significant differences in mean values between inward pressures at both ankle joint positions (p<0.05). The mean TA muscle thickness decreased as inward pressures increased. The significant mean differences between the 1.0 N and 2.0 N, 1.0 N and 4.0 N, 2.0 N and 4.0 N were 0.71, 1.90, and 1.19 respectively at 0° NEU (p<0.05, Figure 3). The significant mean differences between the 1.0 N and 2.0 N, 1.0 N and 4.0 N, 2.0 N and 4.0 N were 0.31, 0.87, and 0.56 respectively at 15° DF (p<0.05, Figure 3).

Discussion

To our knowledge, this is first study to investigate the consistency of TA muscle thickness according to different inward pressures of the transducer at two ankle joint positions for intra- and inter-rater reliability, and the changes of TA muscle thickness according to the inward pressures using the customized probe force holder. Comprehensively, our findings showed that there were high-agreements in all ankle positions per 1.0 N, 2.0 N, and 4.0 N within the examiner and between examiners. Also, with greater inward pressures of the transducer, there were significant decreases in TA muscle thickness.

A previous study [21] including stroke survivors reported that in 0° NEU of the non-paretic side, the ICC values of TA muscle thickness (0.64-0.88) showed lower intra-rater reliability compared to this study (ICCs: 0.92-0.93 and CV: 0.9%-0.18%). The inward pressure of the transducer was adjusted in real-time with a pressure indicator. In contrast to our experimental setting, a study by Cho et al. [21] applied minimal inward pressure of the transducer for reducing the soft area compression. The two divergent methods may have resulted in differences in intra-examiner reliability, and the inward pressure of the transducer within the examiner applied onto stroke survivors might not have been controlled consistently.

Recently, there was an interesting study [22] using a probe force device, which was used to examine the intra-rater reliability of the transverse abdominis by a novice examiner in comparison to the freehand method. This study [22] suggested that use of the device and view of real-time force, inclination or roll would be helpful in obtaining reliable values of abdominal muscle parameters during ultrasound imaging than the freehand method. In the same context, the current study showed a higher intra-rater reliability of obtaining TA muscle thickness using a custom-made holder with a one- axis loadcell through real-time feedback compared to the freehand method. Therefore, it could be interpreted that using a real-time probe force to maintain consistently improved the reliability of TA muscle thickness measurements and diminished the measurement errors.

In the viewpoint on the effect of pressure, as the inward pressure of the transducer increased, there were significant differences in TA muscle thickness. Likewise, a previous study [12] reported the influence of inward pressures of the transducer (0.5 N, 1.0 N, 2.0 N, and 4.0 N) on lateral abdominal muscle thickness. The current study showed a similar trend with the results of a study by Ishida and Watanabe [12]. In regards to the findings of this study, the mean difference (1.90 mm) between 1.0 N and 4.0 N in 0° NEU was greater than MDC in 1.0 N at 0° NEU (1.26 mm). In other words, this indicated that influence of inward pressure of the transducer could interfere with investigating a meaningful change in real TA muscle thickness after dorsiflexor muscle strength training.

However, unlike the results of 0° NEU, the mean difference (0.87 mm) between 1.0 N and 4.0 N in 15° of DF showed lower values than MDC in the condition of 1.0 N at 15° DF (1.11 mm). Namely, in 15° of DF position, it might be permissible for examiners to apply the TA muscle imaging within inward pressure ranges from 1.0 N to 4.0 N. Also, the intra- and inter-rater reliability in 15° DF position were higher than 0° NEU. This may be explained based on a previous mechanical study [23] that TA muscle stiffness may have been accountable for the result when positioned passively at an ankle joint angle. Previous results [23] reported the compression of the crural fascia might stiffen the superficial layer of the TA muscle of some participants as the ankle joint angle was getting approximating into dorsiflexion. This characteristic of the TA muscle could attribute to the easy fixation of the transducer by the examiner. It may explain why the ICCs of TA muscle measurement at 15° DF showed very good reliability and why influence of inward pressure was less than 0° NEU.

The remarkable finding was that as the inward pressure increased, the ICC values of muscle thickness improved. To obtain a stable position, orientation, and consistent inward pressure of the transducer, delicate attention is required during rehabilitative ultrasound imaging [24]. When specific inward pressure is applied, vibration and the gap between the gel and skin might be minimized. It could be interpreted that specific inward pressure is helpful to steadying the position comparatively. However, since the objective of this study was not to determine the optimal inward pressure for measuring the TA muscle thickness using rehabilitative ultrasound imaging, it may be difficult to make generalizations of its use with certain pressures.

The meaningful findings of this study support that when using a custom-made holder with a one-axis load cell and indicator of pressures in real time, TA muscle thickness measurements showed high intra- and inter-rater reliability in different inward pressures at both ankle joint angle. However, there are some limitations in this research. The TA muscle thickness of healthy individuals at passive resting position was assessed, not during contraction. For clinical application, future studies should examine the reliability of TA muscle parameters, such as muscle thickness, pennation angle, and cross-sectional area in persons with stroke or cerebral palsy in regards to impairment of the dorsiflexors utilizing a force indicator in real-time when the TA muscle was contracted. Also, we only measured in limited condition of 1.0 N, 2.0 N, and 4.0 N) in dorsiflexion condition. In the next study, the application of a lower inward pressure (e.g. 0.5 N) is suggested.

The current study highlights the need for trying to view the inward pressure of transducer in real time when measuring TA muscle thickness during rehabilitative ultrasound imaging, particularly at 0° NEU than 15° DF. The muscle has personal characteristics and is likely to be impacted from multivariate factors such as age [25] and sex differences. For this reason, in order to acquire exact change of muscle thickness, the assessor should take initiative to reduce the variability in inward pressures applied through the transducer within the patient into account.

Conflict of Interest

The authors declared no potential conflicts of interest with respect to the authorship and/or publication of this article.

Figures
Fig. 1. (A) The custom-made device with one-axis digital load cell sensor (CSBA-10LS; Curiotech Inc., Seoul, Korea) embedded in the socket. (B) The indicator (CI-2001A; CAS Korea, Seoul, Korea) that checks the measured inward pressure in real time.
Fig. 2. Bland-Altman plots to assess intra-rater reliability of muscle thickness of tibialis anterior. (A) A 0° neutral ankle position (NEU) applied with 1.0 N, (B) 0° NEU applied with 2.0 N, (C) 0° NEU applied with 4.0 N, (D) 15° ankle dorsiflexion (DF) applied with 1.0 N, (E) 15° DF applied with 2.0 N, (F) 15° DF applied with 4.0 N.
Fig. 3. Significant differences of mean values between inward pressures at each ankle position (*p<0.05). NEU: neutral ankle position; DF: ankle dorsiflexion.
Tables

Table 1

Intra-rater reliability for TA muscle thickness in two ankle joint angles at different inward pressures (N=24)

Ankle angleConditionsSession 1Session 2ICC3,3 (95% CI)SEM (mm)MDC (mm)CV (%)
0° NEU1.0 N14.76 (1.56)14.76 (1.61)0.92 (0.82-0.97)0.451.260.13
2.0 N15.05 (1.59)14.05 (1.69)0.92 (0.83-0.97)0.471.290.14
4.0 N12.86 (1.68)12.92 (1.76)0.93 (0.83-0.97)0.481.340.16
15° DF1.0 N15.69 (1.38)15.70 (1.37)0.92 (0.82-0.96)0.401.110.11
2.0 N15.38 (1.39)15.40 (1.43)0.94 (0.86-0.97)0.371.010.10
4.0 N14.82 (1.43)14.87 (1.49)0.96 (0.90-0.98)0.310.860.09

Values are presented as mean (SD) or number only.

TA: tibialis anterior, ICC: intraclass correlation coefficient, CI: confidence interval, SEM: standard error of measurement, MDC: minimal detectable change, CV: coefficient of variation, NEU: neutral ankle position, DF: ankle dorsiflexion.


Table 2

Inter-rater reliability for TA muscle thickness in two ankle joint angles at different inward pressures (N=24)

Ankle angleConditionsExaminer AExaminer BICC2,3 (95% CI)SEM (mm)MDC (mm)CV (%)
0° NEU1.0 N14.76 (1.56)14.75 (1.72)0.89 (0.77-0.95)0.561.560.17
2.0 N15.05 (1.59)14.08 (1.71)0.90 (0.79-0.96)0.511.420.15
4.0 N12.86 (1.68)12.91 (1.77)0.91 (0.80-0.96)0.521.450.17
15° DF1.0 N15.69 (1.38)15.71 (1.41)0.97 (0.92-0.99)0.260.710.07
2.0 N15.38 (1.39)15.41 (1.39)0.97 (0.94-0.99)0.230.640.06
4.0 N14.82 (1.43)14.84 (1.44)0.97 (0.93-0.99)0.250.690.07

Values are presented as mean (SD) or number only.

TA: tibialis anterior, ICC: intraclass correlation coefficient, CI: confidence interval, SEM: standard error of measurement, MDC: minimal detectable change, CV: coefficient of variation, NEU: neutral ankle position; DF: ankle dorsiflexion.


References
  1. Abe T, Loenneke JP, Thiebaud RS. Ultrasound assessment of hamstring muscle size using posterior thigh muscle thickness. Clin Physiol Funct Imaging 2016;36:206-10.
    Pubmed CrossRef
  2. Miyatani M, Kanehisa H, Ito M, Kawakami Y, Fukunaga T. The accuracy of volume estimates using ultrasound muscle thickness measurements in different muscle groups. Eur J Appl Physiol 2004;91:264-72.
    Pubmed CrossRef
  3. Ohata K, Tsuboyama T, Ichihashi N, Minami S. Measurement of muscle thickness as quantitative muscle evaluation for adults with severe cerebral palsy. Phys Ther 2006;86:1231-9.
    Pubmed CrossRef
  4. Thoirs K, English C. Ultrasound measures of muscle thickness: intra-examiner reliability and influence of body position. Clin Physiol Funct Imaging 2009;29:440-6.
    Pubmed CrossRef
  5. Bentman S, O'Sullivan C, Stokes M. Thickness of the middle tra-pezius muscle measured by rehabilitative ultrasound imaging: description of the technique and reliability study. Clin Physiol Funct Imaging 2010;30:426-31.
    Pubmed CrossRef
  6. Cho KH, Lee HJ, Lee WH. Reliability of rehabilitative ultrasound imaging for the medial gastrocnemius muscle in post-stroke patients. Clin Physiol Funct Imaging 2014;34:26-31.
    Pubmed CrossRef
  7. Koppenhaver SL, Hebert JJ, Fritz JM, Parent EC, Teyhen DS, Magel JS. Reliability of rehabilitative ultrasound imaging of the transversus abdominis and lumbar multifidus muscles. Arch Phys Med Rehabil 2009;90:87-94.
    Pubmed CrossRef
  8. O'Sullivan C, Bentman S, Bennett K, Stokes M. Rehabilitative ultrasound imaging of the lower trapezius muscle: technical description and reliability. J Orthop Sports Phys Ther 2007;37:620-6.
    Pubmed CrossRef
  9. Jansen M, van Alfen N, Nijhuis van der Sanden MW, van Dijk JP, Pillen S, de Groot IJ. Quantitative muscle ultrasound is a promising longitudinal follow-up tool in Duchenne muscular dystrophy. Neuromuscul Disord 2012;22:306-17.
    Pubmed CrossRef
  10. Crofts G, Angin S, Mickle KJ, Hill S, Nester CJ. Reliability of ultrasound for measurement of selected foot structures. Gait Posture 2014;39:35-9.
    Pubmed CrossRef
  11. Kristjansson E. Reliability of ultrasonography for the cervical multifidus muscle in asymptomatic and symptomatic subjects. Man Ther 2004;9:83-8.
    Pubmed CrossRef
  12. Ishida H, Watanabe S. Influence of inward pressure of the transducer on lateral abdominal muscle thickness during ultrasound imaging. J Orthop Sports Phys Ther 2012;42:815-8.
    Pubmed CrossRef
  13. Ishida H, Watanabe S. Influence of inward pressure applied by the transducer on trunk muscle thickness during ultrasound imaging. Kawasaki J Med Welf 2014;19:32-7.
  14. Flavell CA, Marshman LG, Gordon SJ. Measurement of trans-versus abdominis activation in chronic low back pain patients using a novel standardized real-time ultrasound imaging method. Ultrasound 2019;27:31-7.
    Pubmed KoreaMed CrossRef
  15. English C, McLennan H, Thoirs K, Coates A, Bernhardt J. Loss of skeletal muscle mass after stroke: a systematic review. Int J Stroke 2010;5:395-402.
    Pubmed CrossRef
  16. Bland DC, Prosser LA, Bellini LA, Alter KE, Damiano DL. Tibialis anterior architecture, strength, and gait in individuals with cerebral palsy. Muscle Nerve 2011;44:509-17.
    Pubmed KoreaMed CrossRef
  17. Kim S, Son J, Yi C, Kim D, Kim Y. Pennation angles of ankle dorsiflexor and plantarflexors depending on muscle contraction intensity. Int J Precis Eng Manuf 2013;14:855-8.
    CrossRef
  18. Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 2016;15:155-63.
    Pubmed KoreaMed CrossRef
  19. Portney LG, Watkins MP. Foundations of clinical research: applications to practice. New Jersey: Pearson/Prentice Hall; 2009.
  20. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;327:307-10.
    Pubmed CrossRef
  21. Cho KH, Lee HJ, Lee WH. Intra- and inter-rater reliabilities of measurement of ultrasound imaging for muscle thickness and pennation angle of tibialis anterior muscle in stroke patients. Top Stroke Rehabil 2017;24:368-73.
    Pubmed CrossRef
  22. Kennedy VL, Flavell CA, Doma K. Intra-rater reliability of transversus abdominis measurement by a novice examiner: comparison of “freehand” to “probe force device” method of real-time ultrasound imaging. Ultrasound 2019;27:156-66.
    CrossRef
  23. Koo TK, Guo JY, Cohen JH, Parker KJ. Quantifying the passive stretching response of human tibialis anterior muscle using shear wave elastography. Clin Biomech (Bristol, Avon) 2014;29:33-9.
    Pubmed CrossRef
  24. Ishida H, Suehiro T, Kurozumi C, Ono K, Watanabe S. Correlation between abdominal muscle thickness and maximal expiratory pressure. J Ultrasound Med 2015;34:2001-5.
    Pubmed CrossRef
  25. Kim MK, Ko YJ, Lee HJ, Ha HG, Lee WH. Ultrasound imaging for age-related differences of lower extremity muscle architecture. Phys Ther Rehabil Sci 2015;4:38-43.
    CrossRef

 

Full Text(PDF) Free

Cited By Articles
  • CrossRef (0)
  • Authorship and ethical issues