Reliability and Validity Study of Inertial Sensor-Based Application for Static Balance Measurement
Phys Ther Rehabil Sci 2022;11:311-20
Published online September 30, 2022
© 2022 Korean Academy of Physical Therapy Rehabilitation Science.

Young Jae Parka , Ho Young Jangb , Kwon Hoi Kimc , Dong Ki Hwangd , Suk Min Leed*

aDivision of Rehabilitation unit, Department of Rehabilitation Medicine, Kangbuk Samsung Hospital, Seoul, Republic of Korea
bDepartment of Physical Therapy, Catholic University Eunpyeong St. Mary’s Hospital
cDepartment of Physical Therapy, Catholic University Uijeongbu St. Mary’s Hosipital
dDepartment of Physical Therapy, Graduate School of Sahmyook University, Seoul, Republic of Korea
Correspondence to: Suk Min Lee (ORCID https://orcid.org/0000-0002-6062-956X)
Department of Physical Therapy, College of Health and Welfare, Sahmyook University, Seoul, Republic of Korea
Tel: +82-10-3704-5650 Fax: +82-2-3399-1639 E-mail: leesm@syu.ac.kr
Received June 19, 2022; Revised August 14, 2022; Accepted September 16, 2022.
cc This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract
Objective: To investigate the reliability and validity of static balance measurements using an acceleration sensor and a gyroscope sensor in smart phone inertial sensors.
Design: Equivalent control group pretest-posttest.
Methods: Subjects were forty five healthy adults aged twenty to fifty-years-old who had no disease that could affect the experiment. After pre-test, all participants wore a waist band with smart phone, and conducted six static balance measurements on the force plate twice for 35 seconds each. To investigate the test-retest reliability of both smart phone inertial sensors, we compared the intra-correlation coefficient (ICC 3, 1) between primary and secondary measurements with the calculated root mean scale-total data. To determine the validity of the two sensors, it was measured simultaneously with force plate, and the comparision was done by Pearson’s correlation.
Results: The test-retest reliability showed excellent correlation for acceleration sensor, and it also showed excellent to good correlation for gyroscope sensor(p<0.05). The concurrent validity of smartphone inertial sensors showed a mostly poor to fair correlation for tandem-stance and one-leg-stance (p<0.05) and unacceptable correlation for the other postures (p>0.05). The gyroscope sensor showed a fair correlation for most of the RMS-Total data, and the other data also showed poor to fair correlation (p<0.05).
Conclusions: The result indicates that both acceleration sensor and gyroscope sensor has good reliability, and that compared to force plate, acceleration sensor has unacceptable or poor correlation, and gyroscope sensor has mostly fair correlation.
Keywords : Telemedicine, Wearable Technology, Postural Balance, Smartphones
Introduction

Fall refers to falling to a lower position or floor than an individual's intention during daily life [1] and falls due to loss of balance can cause serious complications such as trauma and fractures [2]. In particular, aging causes functional loss and degeneration of all organs and tissues. Also, it increases instability in everyday life [3]. According to the Korea Ministry of Health and Welfare survey of senior citizens aged 65 and older, the rate of falling is about 15.9% per year, and the rate of hospitalization is about 64.9% [4].

Balance is the ability to maintain the center of gravity (COG) of the body in the basal plane [5], and is an essential component of functional activity during daily life [2]. The body requires delicate interactions of the motor and sensory systems to maintain balance [6], however, it is reduced by aging-induced muscle strength, impaired proprioceptive sense, vestibular and visual impairment, and delayed response time, as a result, increases the risk of falls [7, 8].

Currently, various examination methods are being tried for balance evaluations. Romberg test can be applied conveniently without any space constraints and is still widely used today as an easy method. Although such a highly utilized evaluation method, it has the disadvantage that quantitative measurement is difficult [9]. Clinics use a lot of equipment to measure center of pressure (COP) in force plate and insole, but it is expensive and requires a large space and expertise [10, 11]. On the other hand, smartphones have less space-time constraints, and frequency collection using acceleration sensors shows high potential for static balance evaluation [12].

Recently, in preparation for the prolonged COVID-19 pandemic, the government has made efforts to make non-face-to-face medical care a regular medical service [13], which has highlighted the importance of tele-rehabilitation. However, there are challenges for clinicians in understanding and technology to evaluate and manage patients remotely [14].

Smartphones are easily accessible and have high potential in tele-rehabilitation for neurological patients, and are actually studying various rehabilitation possibilities using smartphones [15-17]. An inertial measurement unit (IMU), which is key to various rehabilitation studies using smartphones, is a device that measures the speed, direction, gravity, and acceleration of moving objects, consisting of acceleration sensors, geomagnetic sensors, and gyroscope sensors [18]. Various studies have been conducted clinically to develop a smartphone-based balance assessment system by attaching acceleration sensors and gyroscope sensors to the body [17-20].

However, due to the lack of studies that evaluate the reliability of measuring postural sway through two sensors mounted on smartphones while verifying their validity against other balance equipment, it is somewhat difficult to utilize the two sensors in clinical setting.

Therefore, we would like to utilize acceleration sensors and gyroscope sensors to evaluate their balance abilities, verify their reliability and validity, and propose a new balance evaluation methods that can be conveniently used without space-time constraints.

Method

1. Participants

52 healthy adults aged 20 to 50 in the K hospital in Seoul were selected as subjects based on their agreement. Five people who did not meet the pre-experimental criteria were excluded, and two were eliminated during the experiment. Thus, 45 subjects were finally selected. After collecting general features including gender, age, weight, and height. The pre-test was measured with the dominant foot kicking the ball through a ball kick [21] (Table 1). Exclusion criteria were those who had fallen in their daily lives, those who had nervous system, musculoskeletal system disease or damage, and those who had exercised excessively within the past week. This study was reviewed by the Institutional Review Board of Sahmyook University, Seoul. The serial number for the review is 2-1040781-A-N-012021015HR.

2. Measures

The participants of this study wore their smartphones as a belt around their waist (Sacrum 2) and performed the six static balance tests; Shoulder-width-stance with eye open (SWS-EO), Shoulder-width-stance with eye closed (SWS-EC), Feet-together-stance (FTS), Semi-tandem-stance (STS), Tandem-stance (TS), and One-leg stance (OLS) [20, 22] (Figure 1). In order to verify the concurrent validity, all posture measurements were made on the force plate, and simultaneously through an application that records the smartphone inertial sensors. They were repeated twice for 35 seconds. When tests were completed, results were repeated a day later (24 hours) with the same postures and methods in order to identify the test-retest reliability [19]. A Galaxy S20 (SM-G981N, Samsung, Korea) mounted with an inertial sensor (LSM6DSO MEMS, STMicroelectronics, Switzerland) was used to measure the static balance tests. The smartphone was set to horizontal mode and fixed with a belt to the subject's waist (Sacrum 2), and Sensor kinetics pro (Sensor kinetics pro ver.3,1.2, Innovations Inc., US) was run simultaneously with the start of the experiment to measure acceleration data and gyroscope data.

Acceleration data

The acceleration data was based on the Z-axis movement as Accelerometer root mean scale anteroposterior (ARMS-AP) and the Y-axis movement as Accelerometer root mean scale mediolateral (ARMS-ML), and the values obtained through the square root of the Z-axis and Y-axis was normalized to the accelerometer root mean scale total (ARMS-Total).

Gyroscope data

Gyroscope data was the gyroscope root mean scale pitch (GRMS-Pitch) for the movement of the pitch and the gyroscope root mean scale roll (GRMS-Roll) for the movement of the roll, and the gyroscope root mean scale total (GRMS- Total).The data calculated by the gyroscope sensor was multiplied by 180÷π to the calculated data value to change the unit from rad/s to degree/s [20] (Figure 2).

Force plate data

The data of the force plate measured by the movement of COG was normalized to the following three data; Center of pressure velocity anteroposterior (COPV-AP), Center of pressure velocity mediolateral (COPV-ML), and Mean sway velocity (MSV) obtained through the square roots of the X-axis and Y-axis.

3. Data analysis

SPSS software (SPSS ver.20.0, IBM SPSS Statistics, USA) was used for all operations and statistics, and descriptive statistics were used for general characteristics. In order to confirm the reliability of the acceleration data and gyroscope data measured in the first and second trials, each data was analyzed as an intra-class correlation coefficient (ICC 3,1). Pearson's correlation analysis was used to verify the validity of the smartphone inertial sensor-based application. A significance level was set at a p-value of <0.05.

Result

1.Test-retest reliability of smartphone inertial sensors

The reliability of the acceleration sensor showed a very high correlation (ICC>0.90) in all postures (p<0.001). The reliability of the gyroscope sensor showed a very high (ICC>0.90) or high correlation (ICC>0.75) in all postures (p<0.001), which can be shown as below (Table 2).

2.Concurrent validity of smartphone inertial sensors and force plate

Most of tandem-stance and one-leg-stance showed a poor to fair correlation (0.30.05) in a comparison of the acceleration sensor correlation with the force plate (Table 3, Figure 3). However, fair correlation (0.5Table 4, Figure 4).

Discussion

Deterioration in the ability to maintain posture is the cause of falls, and restoration of balance becomes a key goal in the rehabilitation process. Thus, objective evaluation of balance is clinically an important indicator in tracking disease or performing therapeutic interventions [24]. Since the force plate measures the movement of the COP, it measures the velocity on the ground. As activities to increase body stability in both feet increase, the values proportionately increase. On the other hand, the smartphone inertial sensor measures the value of acceleration and angular velocity through the movement of COM from the body. As the body's activity increases, these values become larger [23] Therefore, two signals differ slightly in the movement position, so the relationship cannot be expressed as completely linear [19]. However, because both variables show body sway in proportion to the numerical value and the inertial measurement technologies have the potential to be a reliable alternative compared to the balance evaluation of the force plate [25], many studies have compared simultaneous validity through analysis between the two signals [19, 23, 26, 27]. Therefore, this study tried to verify that the acceleration sensor and gyroscope sensor mounted in smartphones have reliability and validity in measuring the static balance abilities.

First, it was found that both sensors have high reliability when measuring six static balance tests. The studies of Hou et al. [20] and Lee. D. H et al. [28] showed similar results to this study that both sensors have high reliability, and supports the results of this study. On the other hands, the reliability of this study was higher than the study of De Groote et al. [19], which reported that the smartphone acceleration sensor has moderate-high reliability. The reason is that this study used a smartphone equipped with an LSM6DSO sensor with improved accuracy than the LSM6DS3 sensor used in De Grootés research. In addition, the study was analyzed through the RMS values of the Z-axis and Y-axis data, respectively, however, in this study, the mechanical error was corrected by analyzing the total RMS obtained through the square of the two axes, and the reliability value could be increased. Therefore, it is believed that the smartphone acceleration sensor and gyroscope sensor can measure the static balance consistently and reliably.

Second, this study found that it has a higher validity when measuring balance abilities with the gyroscope sensor than with the acceleration sensor. It shows similar results to the studies of De Groote et al. [20] and Han Seul-gi et al. [29] that there was negligible correlation when performing balance tasks with low difficulty level, and that there was a fair or higher correlation when performing balance tasks with higher difficulty level. O'Sullivan et al. [30] reported that acceleration data can be measured in more difficult balance tasks, which supports this study. The reason is that, in the bipodal postures, only minute postural fluctuations occur in the ankle with a wide BOS, so it is not recorded as acceleration data measuring instantaneous postural fluctuations. However, in the tandem stance and one leg stance, the movements of the ML and AP axes appear as large fluctuations due to the narrowed left and right BOS, and are recorded in the acceleration sensor. On the other hand, the reason that the correlation of this study was lower than that of Ruopeng Sun et al. [25] is because the subjects of this study were healthy adults with superior balance ability than those with multiple sclerosis. Thus, it was judged that it was difficult to measure minute postural fluctuations in the easy postures with acceleration data representing strong body movements. A similar study by Koller et al. [31] reported that the acceleration sensor was unable to distinguish between mild multiple sclerosis patients and healthy adults in a low-difficulty static balance postures.

Also, most of the previous static balance measurements only studied the correlation between the acceleration sensor and the force plate [32-34]. However, in this study, the correlation between the gyroscope sensor and the force plate was compared, and it was confirmed that there was poor to fair correlations (0.320], it reported that acceleration data did not correlate with Berg Balance Scale (BBS) (p<0.05), but gyroscope data had strong negative correlations with it (-0.69435] comparing chronic stroke patients, there was no significant correlation between smartphone acceleration sensor and BBS in static balance measurements (p>0.05). However, the data measured with the gyroscope sensor showed a significant correlation with it (p<0.05). These results are consistent with the results of this study that most of the acceleration sensor data had no correlation, but the gyroscope data had poor to moderate correlations. Among the two sensors measured by the smartphone application, the acceleration sensor is easy to measure static balance tests in difficult postures, and the gyroscope sensor is expected to be used to measure most static balance tests. The reason for the difference in correlation between the two devices is that acceleration data can represent strong body movements, whereas gyroscope data represents the degree of how much postural sway appears. As a result, acceleration data is mainly used to analyze the direction and speed of body movement, while gyroscope data analyzes how much the body moves while performing the task in rotational movement and amplitude. Thus, it is necessary to appropriately sum the values obtained through the acceleration sensors and the gyroscope sensors, and correct the error. Therefore, the combination of the two inertial sensors by using additional algorithms such as Kalman filter or Complementary filter could better measure the balance abilities [35]. Park, D, S. et al. [36] reported that there are needs for balance evaluation equipment that can be used immediately and can be evaluated at low cost, instead of existing expensive balance equipment. In this study, we confirmed the possibility of using a smartphone as a method to measure static balance tests as a low cost and easy method in an environment where setting of large equipment is difficult, though measuring the static balance abilities through a smartphone may be less sensitive than measuring the balance using a force plate [19]. Therefore, this study is meaningful as a study that first considered the possibility that a smartphone gyroscope sensor can be used as a new balance evaluation tool in a period of expanding tele-rehabilitation.

There are several limitations in this study. First, this study fixed the smartphone to sacrum two. However, the result values can differ depending on the location of the smartphone attachment because the movements of the ankle, knee, and hip joint may appear as different balance strategies. Therefore, efforts are required to try to quantitatively measure balance tests in various body parts. Second, the sample in the study was homogeneous, consisting of healthy adults aged 20 to 50. In future studies, it will be necessary to conduct studies on various groups, such as patients with difficulty in balance and elderly people with high risk of falling. Third, the test interval was short, so the learning effect was likely to be involved. In future studies, it is necessary to verify the reliability in various aspects. Finally, it took a lot of time to organize a large amount of data because Sensor kinetics pro used to derive inertial data is not a system developed solely to measure balance. In future studies, it will be necessary to develop an application that can calculate data and implement it with an easy quantitative measurement methods.

Figures
Fig. 1. Static Balance Measurement Postures
Fig. 2. Detailed Data Conversion Equations
Fig. 3. Validity of acceleration sensor and force plate
Fig. 4. Validity of gyroscope sensor and force plate
Tables

Table 1

General characteristics of participations (N=45)

Mean±SD
Gender(M/F) 22/23
Age (year) 34.64±8.33
Heights (cm) 167.42±7.73
Weight (kg) 62.27±11.22
Dominant Foot (R/L) 40/5

Table 2

Reliability of smartphone inertial sensors (N=45)

Sensor Postures a ICC b 95% CI
Acceleration SWS-EO 0.978** 0.960-0.988
SWS-EC 0.915** 0.846-0.953
FTS 0.958** 0.924-0.977
STS 0.962** 0.931-0.979
TS 0.916** 0.847-0.953
OLS 0.949** 0.908-0.971
Gyroscope SWS-EO 0.885** 0.791-0.937
SWS-EC 0.902** 0.822-0.946
FTS 0.840** 0.709-0.912
STS 0.786** 0.610-0.882
TS 0.760** 0.564-0.868
OLS 0.903** 0.824-0.947

a Intra-class correlation coefficient (ICC 3,1).

b 95% Confidence Interval.


Table 3

Validity of acceleration sensor and force plate (N=45)

Postures ARMS/FP COPV a PCC (p) b 95% CI
SWS-EO AP 0.013 (0.937) -0.347∼0.533
ML 0.130 (0.419) -0.341∼402
Total/MSV 0.157 (0.328) -0.290~0.569
SWS-EC AP 0.015 (0.549 -0.160∼0.394
ML 0.096 (0.548) -0.160∼0.394
Total/MSV 0.104 (0.516) -0.167∼0.431
FTS AP -0.089 (0.604) -0.261∼0.124
ML 0.143 (0.371) -0.057∼0.419
Total/MSV 0.004 (0.981 -0.160∼0.204
STS AP 0.028 (0.862) -0.245∼0.337
ML 0.102 (0.524) -0.071∼0.464
Total/MSV 0.121 (0.452) -0.092∼0.373
TS AP 0.405 (0.009)** -0.151∼0.719
ML 0.289 (0.067) -0.006∼0.564
Total/MSV 0.408 (0.008)** -0.016∼0.681
OLS AP 0.299 (0.057) -0.101∼0.590
ML 0.556 (0.000)** 0.291∼0.749
Total/MSV 0.542 (0.000)** 0.303∼0.725

a Pearson’s correlation coefficient.

b 95% Confidence Interval.


Table 4

Validity of gyroscope sensor and force plate

Postures GRMS/FP COPV a PCC (p) b 95% CI
SWS-EO Pitch/AP 0.470 (0.009)** 0.068∼0.711
Roll/ML 0.427 (0.019)* 0.015∼0.683
Total/MSV 0.623 (0.000)** 0.277∼0.811
SWS-EC Pitch/AP 0.470 (0.009)** 0.144∼0.734
Roll/ML 0.560 (0.001)** 0.255∼0.771
Total/MSV 0.661 (0.000)** 0.385∼0.839
FTS Pitch/AP 0.709 (0.000)** 0.519∼0.842
Roll/ML 0.413 (0.023)* 0.049∼0.664
Total/MSV 0.684 (0.000)** 0.466∼0.829
STS Pitch/AP 0.645 (0.000)** 0.385∼0.839
Roll/ML 0.474 (0.008)** 0.110∼0.791
Total/MSV 0.623 (0.000)** 0.300∼0.849
TS Pitch/AP 0.364 (0.019)* 0.079∼0.629
Roll/ML 0.636 (0.000)** 0.381∼0.822
Total/MSV 0.450 (0.003)** 0.074∼0.719
OLS Pitch/AP 0.559 (0.001)** 0.358∼0.740
Roll/ML 0.663 (0.000)** 0.443∼0.829
Total/MSV 0.640 (0.000)** 0.440∼0.799

a Pearson’s correlation coefficient.

b 95% Confidence Interval.


References
  1. Nam SM, Na SS, Lee DY. Effects of Otago Exercise program on physical function and risk of falling in the falls experienced elderly. J Korean Acad Nurs. 2020;21(9):424-31.
  2. Park SK, Ryu S, Kim J, Yoon S, Ryu J. Complexity comparison of center of pressure between fallers and non-fallers during gait. KJSB. 2019;29(2):113-9.
  3. Andersen CK, Wittrup-Jensen KU, Lolk A, Andersen K, Kragh-Sørensen P. Ability to perform activities of daily living is the main factor affecting quality of life in patients with dementia. Health Qual. Life Outcomes. 2004;2(1):1-7.
    Pubmed KoreaMed CrossRef
  4. Fall experience of older adults [Internet]. 2018.
    Available from: https://kosis.kr/statHtml/statHtml.do?org Id=117&tblId=DT_117071_050&conn_path=I.
  5. Nichols DS, Miller L, Colby LA, Pease WS. Sitting balance: its relation to function in individuals with hemiparesis. Arch Phys Med Rehabil. 1996;77(9):865-9.
    CrossRef
  6. Horak FB. Postural orientation and equilibrium: what do we need to know about neural control of balance to prevent falls? Age Ageing. 2006;35 (suppl_2):ii7-ii11.
    Pubmed CrossRef
  7. Kim E. The effect of a virtual reality program on static balance control and fall efficacy of elderly people. J Korean Gerontological Soc. 2010;30:1117-6.
  8. Hwang S, Woo Y. Fall prevention strategies in community-dwelling older adults aged 65 or over with type 2 diabetes mellitus: a systematic review and meta-analysis. Phys Ther Rehabil Sci. 2018; 7(4):197-203.
    CrossRef
  9. Yoon S, An H, Ahn T, Choi H, Lee B, Seo D, et al. Performance Evaluation of Balance Ability Equipment Using VR. J Korean soc integr med. 2020;8(3):33-41.
  10. Chang W-D, Chang W-Y, Lee C-L, Feng C-Y. Validity and reliability of wii fit balance board for the assessment of balance of healthy young adults and the elderly. Phys Ther Rehabil Sci. 2013; 25(10):1251-3.
    Pubmed KoreaMed CrossRef
  11. Chesnin KJ, Selby-Silverstein L, Besser MP. Comparison of an in-shoe pressure measurement device to a force plate: concurrent validity of center of pressure measurements. Gait Posture. 2000;12(2): 128-33.
    CrossRef
  12. Hwang J, Hwang S. Smartphone Based Standing Balance Evaluation Using Frequency Domain Analysis of Acceleration. Phys Ther Korea. 2018; 25(3):27-38.
    CrossRef
  13. Kim JY, Lee KI. Advantages and necessities of telehealth care service. Korean J Med. 2020;95(4): 217-27.
    CrossRef
  14. Leochico CFD. Adoption of telerehabilitation in a developing country before and during the COVID-19 pandemic. Ann. Phys Rehabil Med. 2020;63(6):563.
    Pubmed KoreaMed CrossRef
  15. An J, Lee BH. Reliability of joint angle during sit-to-stand movements in persons with stroke using portable gait analysis system based wearable sensors. Phys Ther Rehabil Sci. 2019;8(3):146-51.
    CrossRef
  16. Jung SH, Kim SK, Lee JH, Choi SI, Park DS. Reliability and validity of pelvic mobility measurement using a cushion sensor in healthy adults. Phys Ther Rehabil Sci. 2020;9(2):74-81.
    CrossRef
  17. Ferreira CFR. Smartphone Based Tele-Rehabilitation. 2013.
  18. Kim JH, Kim BS, Jeon HM, Kang SY. An LED positioning method using image sensor of a smart device. J-KICS. 2015;40(2):390-6.
    CrossRef
  19. De Groote F, Vandevyvere S, Vanhevel F, De Xivry J-JO. Validation of a smartphone embedded inertial measurement unit for measuring postural stability in older adults. Gait Posture. 2021;84:17-23.
    Pubmed CrossRef
  20. Hou YR, Chiu YL, Chiang SL, Chen HY, Sung WH. Development of a smartphone-based balance assessment system for subjects with stroke. J. Sens. 2019;20(1):88.
    Pubmed KoreaMed CrossRef
  21. Gabbard C, Hart S. A question of foot dominance. J. Gen. Psychol. 1996;123(4):289-96.
    Pubmed CrossRef
  22. Nejc, S., Jernej, R., Loefler, S., Kern, H. Sensitivity of body sway parameters during quiet standing to manipulation of support surface size. Journal of Sports Science & Medicine, 9(3), 431.
  23. Lee CH, Sun TL. Evaluation of postural stability based on a force plate and inertial sensor during static balance measurements. J Physiol Anthropol. 2018;37(1):1-16.
    Pubmed KoreaMed CrossRef
  24. Prosperini L, Fortuna D, Giannì C, Leonardi L, Marchetti MR, Pozzilli C. Home-based balance training using the Wii balance board: a randomized, crossover pilot study in multiple sclerosis. Neurorehabil Neural Repair 2013;27(6):516-25.
    Pubmed CrossRef
  25. Sun R, Moon Y, McGinnis RS, Seagers K, Motl RW, Sheth N, et al. Assessment of postural sway in individuals with multiple sclerosis using a novel wearable inertial sensor. Digit. Biomark. 2018;2(1): 1-10.
    Pubmed KoreaMed CrossRef
  26. Chung, C. C., Soangra, R., Lockhart, T. E. Recurrence quantitative analysis of postural sway using force plate and smartphone. In Proceedings of the Human Factors and Ergonomics Society Annual Meeting. Sage CA: Los Angeles, CA: SAGE Publications. 2014;1271-1275.
    CrossRef
  27. Jebelli, H., Yang, K., Khalili, M. M., Ahn, C. R., Stentz, T. Assessing the effects of tool-loading formation on construction workers’ postural stability. In Construction Research Congress. 2018;292-302.
    CrossRef
  28. H. LD, K. HS, J. JH. A study on the reliability of balance abilities measure using a smartphone. J Neurother. 2020;24(2):1-5.
  29. Han SK, Lee IH, Park NR. Reliability of static balance abilities measure using a smartphone's acceleration sensor. J Korean Acad Nurs. 2016;17(6): 233-8.
    CrossRef
  30. O’sullivan M, Blake C, Cunningham C, Boyle G, Finucane C. Correlation of accelerometry with clinical balance tests in older fallers and non-fallers. Age Ageing. 2009;38(3):308-13.
    Pubmed CrossRef
  31. Koller JR, Jacobs DA, Ferris DP, Remy CD. Learning to walk with an adaptive gain proportional myoelectric controller for a robotic ankle exoskeleton. J NeuroEng Rehabil. 2015;12(1):1-14.
    Pubmed KoreaMed CrossRef
  32. Mellone S, Tacconi C, Chiari L. Validity of a Smartphone-based instrumented Timed Up and Go. Gait Posture. 2012;36(1):163-5.
    Pubmed CrossRef
  33. Lee BC, Kim J, Chen S, Sienko KH. Cell phone based balance trainer. J NeuroEng Rehabil. 2012; 9(1):1-14.
    Pubmed KoreaMed CrossRef
  34. Jeremy AP, Amick RZ, Thummar T, Rogers ME. Validation of measures from the smartphone sway balance application: a pilot study. Int J Sports Phys Ther. 2014;9(2):135.
  35. Hou YR, Chiu YL, Chiang SL, Chen HY, Sung WH. Feasibility of a smartphone-based balance assessment system for subjects with chronic stroke. Comput Methods Programs Biomed. 2018;161:191-5.
    Pubmed CrossRef
  36. Park DS, Lee DY, Choi SJ, Shin WS. Reliability and validity of the balancia using wii balance board for assessment of balance with stroke patients. J Korean Acad Nurs. 2013;14(6):2767-72.